Model Number MMT-751LNAS |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problems
Fatigue (1849); Hyperglycemia (1905); Nausea (1970); Vomiting (2144)
|
Event Date 08/22/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
Currently, it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that she was hospitalized for a high blood glucose of 1,013 mg/dl.The customer experienced nausea, sluggishness, vomiting, and sleeplessness.Her blood glucose began to climb once she programmed her new insulin pump.The customer's settings were reviewed and they were correct.A high pressure test was performed twice and failed.The insulin pump was returned and replaced.
|
|
Manufacturer Narrative
|
The insulin pump passed all functional testing including the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test.No cosmetic damage noted.
|
|
Manufacturer Narrative
|
The pump passed the delivery accuracy test.
|
|