• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; OZO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; OZO Back to Search Results
Model Number MMT-751LNAS
Device Problem Device Operates Differently Than Expected (2913)
Patient Problems Fatigue (1849); Hyperglycemia (1905); Nausea (1970); Vomiting (2144)
Event Date 08/22/2016
Event Type  Injury  
Manufacturer Narrative
Currently, it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that she was hospitalized for a high blood glucose of 1,013 mg/dl.The customer experienced nausea, sluggishness, vomiting, and sleeplessness.Her blood glucose began to climb once she programmed her new insulin pump.The customer's settings were reviewed and they were correct.A high pressure test was performed twice and failed.The insulin pump was returned and replaced.
 
Manufacturer Narrative
The insulin pump passed all functional testing including the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test.No cosmetic damage noted.
 
Manufacturer Narrative
The pump passed the delivery accuracy test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP
Type of Device
OZO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key5950249
MDR Text Key54695260
Report Number3004209178-2016-78828
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
PMA/PMN Number
120010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup,Followup
Report Date 03/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-751LNAS
Device Catalogue NumberMMT-751LNAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received03/29/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age55 YR
Patient Weight77
-
-