Model Number MLE-28-012 |
Device Problems
Break (1069); Material Protrusion/Extrusion (2979)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/01/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
At the time of this report, the device has not yet been returned for evaluation.As a result, a determination cannot be made at this time.If further information becomes available, gyrus acmi will continue the investigation and update the agency accordingly.No device returned.
|
|
Event Description
|
During a turp procedure not long after the start of the procedure the loop appeared to be "burnt out" and the wire broke right at the center/apex of the curve on the loop.As this procedure is performed under direct vision, the surgeon was able to immediately stop the procedure and the broken electrode was switched for another from the same box.The same fault then occurred with the 2nd electrode.This electrode was then switched for an electrode from a new box (with a different lot number) and there were no further problems.The procedure was completed normally with no detrimental effects on the patient.
|
|
Manufacturer Narrative
|
Six unopened electrodes were returned, lot 789563jf.Two electrodes were tested on steak tissue, connected to a valley lab force ii generator set at pure cut 20.The tissue was placed onto a grounding pad, the generator was activated and the cut function applied.Both devices cut the sample tissue, no burn out or loop breakage observed.Could not confirm the customers complaint.No information was provided as to the generator power setting.
|
|
Search Alerts/Recalls
|