Model Number 809810
Device Problems Device Alarm System (1012); Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 08/27/2016
Event Type  Malfunction  
Manufacturer Narrative

(b)(4). The reported complaint was confirmed. The unit has been empty/unused since last preventive maintenance (pm) february 2016. The fsr reseated channel a water level sensor connections at the sensors and the direct current (d/c) control board. The issue resolved. The unit operated to manufacturer specifications and was returned to clinical use. No part will be returned to the manufacturer. If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.

 
Event Description

The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the channel a continuously alarmed ¿low water¿ on the cooler heater unit despite the fact the unit was full of water and primed. There was no patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHX2 TEMPERATURE MANAGEMENT SYSTEM
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-HX2 TEMPERATURE MANAGEMENT SYS.
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5962953
MDR Text Key55443674
Report Number1828100-2016-00631
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK071521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE
Reporter Occupation
Type of Report Initial
Report Date 09/20/2016
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received09/20/2016
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator OTHER
Device MODEL Number809810
Device Catalogue Number809810
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/27/2016
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured09/18/2008
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-