• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CODMAN & SHURTLEFF, INC. / MEDOS S.A. HAKIM PROGRAMMABLE VALVE; SHUNT, CENTRAL NERVOUS SYSTEM & COMPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN & SHURTLEFF, INC. / MEDOS S.A. HAKIM PROGRAMMABLE VALVE; SHUNT, CENTRAL NERVOUS SYSTEM & COMPS Back to Search Results
Catalog Number 82-3113
Device Problems Break (1069); Device Dislodged or Dislocated (2923)
Patient Problem No Information (3190)
Event Date 08/30/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).It has been communicated that the device and/or lot information is not available for evaluation.Without the device and/or lot information it is not possible for codman to conduct a proper investigation.If at some point the device and/or lot information does become available, this complaint will be re-opened, evaluated and a follow up report will be filed.Trends will be monitored for this and similar complaints.At the present time this complaint is considered closed.
 
Event Description
The device was implanted via v-p shunt to a (b)(6) patient 2 years ago.The initial pressure setting is unknown.At an x-ray examination 2 years after implantation, the cam of the valve was found dislodged.Although the ventricles of the patient's brain are enlarged, no other symptoms are noted at this point.So the patient is currently being monitored without a revision.According to the surgeon, the patient may have hit the head, resulting in the valve breakage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAKIM PROGRAMMABLE VALVE
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM & COMPS
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
rue girardet 29
ch-2400 le locle
SZ 
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key5963275
MDR Text Key55137088
Report Number1226348-2016-10675
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K980778
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number82-3113
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/01/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age7 YR
-
-