• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVAMEX TRANSPORT 20 IN X 16 IN 9153637574; WHEELCHAIR, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVAMEX TRANSPORT 20 IN X 16 IN 9153637574; WHEELCHAIR, MECHANICAL Back to Search Results
Model Number TREX2
Device Problems Bent (1059); Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available, a supplemental record will be filed.
 
Event Description
Per (b)(6), she stated that her customer alleged that they just received this wheelchair and there was some damage to the frame.The bar on the left side of the chair if you're sitting in it, this bar fits into the h-hook.The bar itself is bent and won't go into the slot it should fit into.
 
Manufacturer Narrative
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the right seat rail was bent and would not fit the h block, which confirmed the original complaint issue.The underlying cause was freight damage.
 
Event Description
Per lynne at direct supply she stated that her customer alleged that they just received this wheelchair and there was some damage to the frame.The bar on the left side of the chair if you're sitting in it, this bar fits into the h-hook.The bar itself is bent and won't go into the slot it should fit into.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSPORT 20 IN X 16 IN 9153637574
Type of Device
WHEELCHAIR, MECHANICAL
Manufacturer (Section D)
INVAMEX
parque industrial manimex
reynosa 88780
MX  88780
Manufacturer (Section G)
INVAMEX
parque industrial manimex
reynosa 88780
MX   88780
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5965989
MDR Text Key55217895
Report Number9616091-2016-01248
Device Sequence Number1
Product Code IOR
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 09/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberTREX2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received09/30/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-