• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION HOLLOW FIBER OXY WITH 3000ML BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION HOLLOW FIBER OXY WITH 3000ML BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*RX15RE30
Device Problem Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 08/30/2016
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed. Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available. (b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, while the product was sitting on the shelf at the hospital, it noticed that there was damage to the packaging as well as damage to the product inside the packaging. There were 2 holes in the packaging, there was a crushed luer cap at the bottom of the package and one of the cardiotomy luer ports were damaged. No patient involvement as this occurred during out of box.
 
Manufacturer Narrative
This follow-up report is submitted to the fda in accord with applicable regulations - and as indicated by terumo cardiovascular system in the initial report submitted to the fda on september 22, 2016. (b)(4). The returned sample was visually inspected upon receipt. It was found that the packaging, breather bag, and luer port on the top of the reservoir lid were all damaged. The cap on the luer port was broken off of the port. A retention sample from the same product code/lot number combination was visually inspected and confirmed to have no damages. All capiox units are 100% visually inspected at several points in the production process. A review of the device history record revealed no manufacturing anomalies. (b)(4) - human factors issue was chosen because it is likely that the packaging and product were damaged by a shock force at some point during the handling of the product; however, when or how this shock force was applied was not able to be determined. All available information has been placed on file in quality management for appropriate tracking, trending and follow up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHOLLOW FIBER OXY WITH 3000ML
Type of DeviceBLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball rd
elkton MD 21921
Manufacturer Contact
shari bailey
125 blue ball rd
elkton, MD 21921
8002837866
MDR Report Key5970428
MDR Text Key55440355
Report Number1124841-2016-00335
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130333
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/12/2016
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/31/2019
Device Model Number3CX*RX15RE30
Device Catalogue NumberN/A
Device Lot NumberUC29
Other Device ID Number(01)00699753450134
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-