The customer complained of erroneous results for 1 neonate patient tested for bilirubin total gen.3 (bilt3).The erroneous result was reported outside of the laboratory.The initial bilt3 result was 24.6 mg/dl.This result was reported to the doctor.The doctor called back almost immediately and requested the sample be rerun.The sample was repeated twice with results of 15.7 mg/dl and 15.6 mg/dl.The repeat results were believed to be correct.No adverse event occurred.The bilt3 reagent lot number was 13902101 with an expiration date of 09/30/2017.The field service representative ran instrument mechanical and fluid checks and all were acceptable.Precision tests were acceptable.It was noted that quality controls (qc) with multiple lot numbers had been acceptable for the past month.A specific root cause could not be identified.Both instrument and reagent problems can be excluded since calibration and qc were acceptable and the instrument tests were within specification.The most likely root cause of the issue is a pre-analytical issue.It was suggested that the sample may not have been handled correctly.Based on the information provided for investigation, there may have been too much sample volume pipetted for the initial run or some material found its way into the measuring cuvette which caused the higher value.
|