Age at time of event: 18 years or older.Device is a combination product. (b)(4).Device evaluated by mfr: returned device consisted of a coyote es balloon catheter with a stopcock.The distal end of the device (tip, balloon, markerbands, and distal shaft) was not returned for analysis.The hypotube and proximal shaft were tactile and microscopically inspected.Inspection revealed numerous kinks in the hypotube, with a complete separation in the proximal shaft.The separated end appeared to have had excessive tensile force applied to it.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Functional testing could not be done, due to the condition of the returned device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
It was reported that shaft break and catheter entrapment occurred.After a non-bsc guidewire crossed the lesion, a 2mm x 40mm x 145cm coyote es balloon catheter was attempted to be advanced over the guide wire to dilate the lesion.However, the device became stuck with the wire outside the patient and when the physician forcefully pulled the device out, the balloon shaft got broken.The procedure was completed with a non-bsc balloon catheter.No patient complications were reported and the patient's status was good.
|