• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND SORIN CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-01-04
Device Problem Failure to Power Up (1476)
Patient Problem No Patient Involvement (2645)
Event Date 08/30/2016
Event Type  No Answer Provided  
Manufacturer Narrative
There was no patient involvement.Sorin group (b)(4) manufactures the sorin centrifugal pump 5 (cp5).The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the motor of the sorin centrifugal pump 5 (cp5) would not boot during priming.There was no patient involvement.A sorin group field service representative was dispatched to the facility to investigate.The service representative read the machine configuration and event storage of the unit and did not identify any issues.The firmware of the device was updated.The customer reported that the unit had run flawlessly during a test run and on a patient before the service visit.The unit is back in regular use.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the motor of the sorin centrifugal pump 5 (cp5) would not boot during priming.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w 65th.way
arvada, CO 80004
3034676461
MDR Report Key5978829
MDR Text Key55685700
Report Number9611109-2016-00620
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 08/31/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-01-04
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/31/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/20/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-