Catalog Number 532.110VET
Device Problems Electrical /Electronic Property Problem (1198); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/15/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported from (b)(4) that during a veterinary tplo surgery on a canine, it was observed that the small battery drive device motor was slow to rev up or didn't engage at all.It was reported that there were no delays and they were able to complete the surgery using the same device successfully.There was no human patient involvement as this was a veterinary procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device failed the power test at the test bench, coupling head was worn, unknown substance found on internal components, electric motor emits an unusual noise and the wire insulation on the electronic control unit was damaged.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE II VET SALE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5980641
MDR Text Key56174092
Report Number8030965-2016-15007
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.110VET
Device Lot Number1424
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received10/25/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N