Model Number 809810 |
Device Problems
Loose or Intermittent Connection (1371); Device Inoperable (1663)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/07/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The fsr tagged the unit "defective" and has removed it from service.The part is currently unavailable at the manufacturer.
|
|
Event Description
|
It was reported that during notice of field correction (nfc), during release testing of the heater cooler the resistance temperature detector (rtd) connector on the back of the direct current (d/c) control circuit board as loose, and then broke off.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.No part was replaced for this issue, therefore no part will be returned to the manufacturer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|