Model Number MERGE HEMODYNAMICS 10.0.1
Device Problem Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/30/2016
Event Type  malfunction  
Manufacturer Narrative
The faulty ups was returned to merge healthcare on 21sep2016 for evaluation.The results showed that no problem could be found.The unit was setup to run for many days and no further issues occurred.Device labeling, hemo user manual, addresses the potential for such an occurrence with statements such as, "if low voltage is detected, the pdm goes to battery power and an audible alarm sounds.".
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that the ups (uninterrupted power supply) device was faulty and caused the hemo application to go down twice both times during procedures causing the hemo monitor to be rebooted.This resulted in the loss of patient data.With merge hemodynamics not capturing physiological data, there is a potential for incorrect treatment that could cause harm to a patient.However, the user confirmed that, once the hemo monitor was rebooted, the procedures were completed successfully.Subsequently, the user plugged the hemo system into the site's emergency power supply after the second occurrence.This prevented the problem from occurring again while waiting for replacement hardware.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key5989388
MDR Text Key55999098
Report Number2183926-2016-00743
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 10.0.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received08/30/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/21/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-