Catalog Number 700-7177-001
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Test Result (2695)
Event Date 09/08/2016
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) was at the customer site and observed that there was excessive air in the color gravity module (cgm) tubing.Fse also found leaking tubing at the connection to the cgm and replaced the tubing.He ran reflectance check, ran sg color gravity check and ran samples with no further issues.The repairs were verified per established service procedures.Bec internal identifier for this report is (b)(6).
 
Event Description
The customer stated that 4 samples run on their ichem velocity automated urine chemistry system gave an erroneous blue color result.Erroneous patient results were not reported out and there was no change or effect to patient treatment in connection to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM
Type of Device
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer (Section G)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer Contact
gopal mohanty
9172 eton avenue
chatsworth, ca, CA 91311
8185277379
MDR Report Key5990537
MDR Text Key56038991
Report Number2023446-2016-00377
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Catalogue Number700-7177-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-