Model Number 25-40-45
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 09/09/2016
Event Type  malfunction  
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 gas blender system. The incident occured in (b)(6). This medwatch report is being filed on behalf os sorin group (b)(4). Sorin group (b)(4) received a report that a s5 gas blender system displayed an error message during priming. There was no patient involvement. The device has been requested for return to sorin group deutschland for investigation. A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that a s5 gas blender system displayed an error message during priming. There was no patient involvement.
 
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 gas blender system. The incident occurred in (b)(6). This medwatch report is being filed on behalf of livanova (b)(4). The reported device was returned to livanova (b)(4) for further investigation. The described error was reproduced during testing and was traced to a defective measurement bridge (awm 5102vn), air and o2 mass flow controller and a defective fan. These components were replaced to resolve the reported failure. The device was disinfected and cleaned. A test run was performed and no additional failures were noted. An exact root cause for the defective components was not determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameS5 GAS BLENDER SYSTEM
Type of DeviceGAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5998794
MDR Text Key56449830
Report Number9611109-2016-00666
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/12/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number25-40-45
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-