|
Model Number A22201C |
Device Problem
Arcing of Electrodes (2289)
|
Patient Problem
Burn, Thermal (2530)
|
Event Date 09/29/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
The device has not been returned to olympus for evaluation.The cause of the reported event could not be determined.The instruction manual states, ¿improper use of hf current can cause endogenous or exogenous burns, and explosions.Thoroughly review the hf current safety information included in the olympus endoscopy system guide.If any damage to the insulation at the electrode¿s distal end becomes visible during use, replace the electrode with an undamaged electrode.Otherwise there is a risk of injury for the patient and/or user.¿.
|
|
Event Description
|
Olympus was informed that during an unspecified therapeutic procedure, the electrode loop arced and burned the patient's uterus.The insulation at the tip of the loop was also burnt.The procedure was completed using a different electrode loop.It is unknown if the patient required an intervention.
|
|
Manufacturer Narrative
|
Olympus received additional information on october 7, 2016 stating that the reported patient burn occurred during a hysteroscopy and polypectomy procedure when the physician was attempting to remove a polyp from the patient's uterus.It was also reported that the patient's burn was not severe.No medical or surgical intervention was required.The patient is reportedly doing fine.The device was inspected prior to use.
|
|
Manufacturer Narrative
|
The reported device was returned to olympus for evaluation; however, the user facility¿s hysterescope/resectoscope that was used in conjunction with the reported device was not returned to olympus.The evaluation confirmed the reported device damage.A visual inspection of the device noted deformation on both of the resector post.A microscope was used and found charred stains on the metallic post.To add, one of the yellow posts was completely missing.Scrape marks were also noted on the metallic post where the charred stains were located.The scrape marks can occur when the device comes in contact with a hard like/metal objects.Based on the investigation findings, the cause of the reported event is likely attributed to the operator's technique and the device coming in contact with a hard like/metallic object when the high frequency (hf) output was activated, which can cause premature melting of the loop.
|
|
Search Alerts/Recalls
|
|
|