Model Number 4FC12 |
Device Problem
Air Leak (1008)
|
Patient Problems
Air Embolism (1697); ST Segment Elevation (2059)
|
Event Date 09/09/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure air introduction into the patient's left atrium was alleged.Procedure outcome was unknown at this time.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
The sheath was returned and analyzed.Visual inspection of the sheath showed that the shaft was intact with no apparent issues.The reported air ingress could not be reproduced.Multiple aspirations and injections were performed without air bubbles or leaks.Also, the hemostatic valve and the valve assembly were leak tight.In conclusion, the reported air ingress issue was not confirmed through testing.A clinical issue of st elevation was encountered during the procedure.The sheath passed the returned product inspection as per specification.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that air was not visible and "just the st elevation was observed on the ecg that took some time to resolve".No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|