(b)(4).Age at time of event: 18 years or older.Device evaluated by mfr: returned product consisted of a maverick 2 balloon catheter in two pieces.The balloon was tightly folded and there was blood in the lumen.The hypotube and inner/outer shaft was microscopically examined.Inspection revealed a kink in the hypotube 86cm from the strain relief with a separation in the hypotube 88cm from the strain relief.The fracture faces were oval as if kinked prior to separation.There was no evidence of any material or manufacturing deficiencies contributing to the damage.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
|
Reportable based on device analysis completed on 15-sep-2016.It was reported that shaft kink occurred.The 85% stenosed, 28mm in length target lesion was located in the moderately tortuous, moderately calcified, and 3.00mm in diameter left anterior descending artery.A 2.50mm x 20mm maverick²¿ balloon catheter was advanced to treat the lesion.However, the shaft was noted to be kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed hypotube break.It was further reported that during advancing of the device, the shaft broke.
|