• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 ARCTIC SUN GEL PADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 ARCTIC SUN GEL PADS Back to Search Results
Catalog Number 317-02
Device Problem Insufficient Information (3190)
Patient Problem Burn, Thermal (2530)
Event Date 09/19/2016
Event Type  Injury  
Manufacturer Narrative
The device was not returned for evaluation. The lot number is unknown; therefore, the device history record could not be reviewed. The instructions for use states the following: ¿do not place arcticgel pads on skin that has signs of ulcerations, burns, hives or rash. While there are no known allergies to hydrogel materials, caution should be exercised with any patient with a history of skin allergies or sensitivities. Due to underlying medical or physiological conditions, some patients are more susceptible to skin damage from pressure and heat or cold. Patients at risk include those with poor tissue perfusion or poor skin integrity due to diabetes, peripheral vascular disease, poor nutritional status or steroid or high dose vasopressor therapy. If accessible, examine the patient¿s skin under the arcticgel pads often; especially those patients at higher risk of skin injury. Skin injury may occur as a cumulative result of pressure, time and temperature. Do not place bean bags or other firm positioning devices under the arcticgel pads. Do not place any positioning devices under the pad manifolds or patient lines. ¿ (b)(4). The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard. The device was not returned.
 
Event Description
It was reported that a patient with immersion syndrome being treated with the arctic sun device allegedly experienced a burn on the scapular area two days after treatment began. Skin care was provided to the affected area.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN GEL PADS
Type of DeviceARCTIC SUN GEL PADS
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX 85621
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX 85621
Manufacturer Contact
janna parks
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6026302
MDR Text Key57308872
Report Number1018233-2016-01390
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K142702
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation
Type of Report Initial
Report Date 09/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue Number317-02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/21/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/13/2016 Patient Sequence Number: 1
-
-