Model Number N/A |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/16/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Report two of two for the same event, reference 0001032347-2016-00597.
|
|
Event Description
|
The sales associate reported two right angle drivers in which the gears were slipping causing the blade not to spin, or the mechanism for releasing the blade/fixation pin would not release leaving the driver attached to the fixation pin.Additional instruments were used to complete the surgery.No surgical delay or adverse events were reported.
|
|
Manufacturer Narrative
|
The product was returned for evaluation.The product identity was confirmed in the evaluation.The device history records were reviewed in the evaluation and no non-conformances were found.The driver was visually inspected and show signs of normal use.The driver was functionally tested and was found not to function as intended because when load was applied to the knob, it would spin freely without turning the blade or inserting the screw.The driver did pass the max gage inspection.The complaint is confirmed as the driver did not function as intended.The most-likely cause was determined to be excessive torque applied to the driver.This is report two of two for the same event.Report one of two is reported on mfr #0001032347-2016-00597.
|
|
Manufacturer Narrative
|
This is supplemental report two of two for the same event, reference 0001032347-2016-00597-2.
|
|
Search Alerts/Recalls
|