• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS HEART START XL; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS HEART START XL; DEFIBRILLATOR Back to Search Results
Model Number M4735A
Device Problem Defibrillation/Stimulation Problem (1573)
Patient Problem Complaint, Ill-Defined (2331)
Event Date 10/17/2016
Event Type  Injury  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
It was reported to philips that the heartstart xl defibrillator did not shock when needed.Another defibrillator was used to provide therapy to the patient.The patient was taken to intensive care and intubated.
 
Manufacturer Narrative
A philips field service engineer evaluated the device and was unable to duplicate the failure.There were no electronic event files or ecg monitoring waveforms available for review.No parts were replaced.The device passed all performance assurance tests and was placed back into use with the customer.Philips was not able to confirm the reported malfunction.Because the problem could not be recreated, the cause cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART START XL
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
betty harris
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key6040519
MDR Text Key57851942
Report Number1218950-2016-06586
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K001725
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM4735A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/17/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-