• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752037
Device Problem Insufficient Information (3190)
Patient Problems Intraocular Pressure, Delayed, Uncontrolled (1936); Perforation (2001)
Event Date 10/05/2016
Event Type  Injury  
Manufacturer Narrative
No sample has been returned for evaluation; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, lot history and complaint history reviews could not be conducted.Due to an observed increased trend for this event, an internal investigation was initiated to determine if corrective and preventative actions were necessary.A root cause for the increased complaint rate was found to be related to a manufacturing post assembly inspection practice.This complaint does not identify a reported lot and no sample was returned for evaluation so it cannot be determined if the complaint can be attributed to the post assembly inspection.The post assembly inspection has been discontinued and an effectiveness check has been established and will be monitored periodically.(b)(4).
 
Event Description
An ophthalmic surgeon reported that when the trocar cannula was removed from the scleral at the end of a vitreous procedure, a hypotony occurred because of a "poor self-sealing perforation".An another additional procedure was performed.The sample was not available because it was discarded.Additional information was requested; however, none has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key6052370
MDR Text Key58243729
Report Number1644019-2016-01353
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 10/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752037
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/05/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM
Patient Outcome(s) Other; Required Intervention;
-
-