MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.010
Device Problem Failure to Shut Off (2939)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/12/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during sterile processing, it was observed that the small battery drive device would stay engaged even after the trigger was released and the device was turned off.According to the reporter, the only way to stop the device was to remove the battery device.There were no delays to a surgical procedure.It was unknown if a spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device failed the start stop test, stays running on its own, the coupling head worn, there was an unknown residue/substance found on the internal components, the electric motor emitted an unusual noise and the ball bearings seized.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper cleaning/care and component wear from use and servicing over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6060212
MDR Text Key59013017
Report Number8030965-2016-15362
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.010
Device Lot Number3793
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received11/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-