Model Number 24501879
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/29/2016
Event Type  malfunction  
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and functioned at low suction.Dark grey substance, e.G battery acid, was observed inside the battery compartment.Internally, no signs of foreign substance, burning, melting or charring were observed.The returned batteries were observably damaged.Additional testing confirmed that batteries may leak when the upper middle battery is placed incorrectly, with the positive and negative end reversed.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2016 to report the pump base of the purely yours ultra breast pump became hot and batteries began leaking black sticky fluid while pumping on (b)(6) 2016.She was not using the ac adapter with batteries installed in the battery compartment.She states coming in contact with the black leaking fluid but denies burn or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS ULTRA
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA, INC.
485 half day rd.
suite 320
buffalo grove IL 60089
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key6063514
MDR Text Key58771267
Report Number3009974348-2016-00230
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial
Report Date 09/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number24501879
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received09/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/07/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age36 YR
-
-