The manufacturer representative (rep) reported that it was taking too long to recharge.
It was reported that the recharge stats from the clinician programmer reported a 2.
9, 6 days, average of 4 coupling bars.
There was a report that repositioning the antenna resolved the issue.
The rep was able to get 8 coupling bars and troubleshooting resolved the issue.
It was reported that the rep would work with the patient on education on coupling.
No symptoms were reported.
Relevant medical history includes spinal pain.
On (b)(6) 2016,the patient reported they are having some charging issues with their device.
They stated that they have to charge all day to get their implant fully charged, but they can¿t leave it to charge all day because the recharger shows an overheating warning message.
The patient further stated that their implantable neurostimulator (ins) keeps ¿shorting out.
¿ they further explained that when they take a shower, the ins battery depletes.
It was noted that the patient has been charging every 12 hours for about the past 3 months.
The patient mentioned their therapy hurts when they have it on.
They stated they had a fall in (b)(6) 2016, but x-rays were taken of the leads and nothing unusual was found.
Additionally, the patient cannot turn their neck due to ¿anchors¿ that are in their neck.
The patient also mentioned that they are on so much medication that it is making them sick.
The patient was to follow up with a manufacturing representative to have their device checked.
The patient was implanted for spinal pain.
|