DEPUY SYNTHES POWER TOOLS COLIBRI II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.101 |
Device Problems
Device Stops Intermittently (1599); Battery Problem (2885)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and observed that the trigger was broken and torn off.Therefore the reported condition was duplicated and confirmed.The assignable root cause was determined to be due to normal wear from use and servicing over time.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was discovered that the motor on the small battery drive device stopped working but resumed working after a while.It was unknown if there were any delays to the planned surgical procedure or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|