• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIS INTERNATIONAL IQ 200 SYSTEM; AUTOMATED URINALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIS INTERNATIONAL IQ 200 SYSTEM; AUTOMATED URINALYSIS SYSTEM Back to Search Results
Catalog Number 700-3375
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2016
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) was at the customer site and observed that the tubing from the probe bypass valve (pbv) to the filter was disconnected at the pbv.The green fitting had been broken off and half of the fitting was still inside the pbv.The fse replaced the pbv and focus and then determined the calibration needed to be completed.The fse completed calibration and was able to run positive and negative controls successfully.The repairs were verified as per service procedures.(b)(4).
 
Event Description
The customer reported positive quality control failing low on their iq 200 system.There were no erroneous patient results generated.There was no change to patient treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IQ 200 SYSTEM
Type of Device
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer (Section G)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer Contact
gopal mohanty
9172 eton avenue
chatsworth, ca, CA 91311
8185277379
MDR Report Key6074869
MDR Text Key59035026
Report Number2023446-2016-00394
Device Sequence Number1
Product Code KQO
UDI-Device Identifier10837461001324
UDI-Public(01)10837461001324
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022774
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Other Health Care Professional
Device Catalogue Number700-3375
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-