Catalog Number ASKU
Device Problems Fluid/Blood Leak (1250); Improper or Incorrect Procedure or Method (2017); Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/10/2016
Event Type  malfunction  
Manufacturer Narrative
This is a report of a use error where the final bag fell resulting in a tear in the tubing of a cassette.Use errors and proper user instructions are addressed in ¿the homechoice and homechoice pro systems patient at-home guide¿, which is shipped with every homechoice device.The guide instructs the user to place the solution bags on a flat, stable surface and that falling bags can result in a disconnection or leak.It also provides step-by-step instructions for connecting the solution bags.A review of the label for the product family will be conducted.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a leak occurred on a homechoice cassette.This event occurred during dwell five of five while the patient was connected.The technical service representative (tsr) helped the registered nurse (rn) to end therapy early.The rn had come into the room to find the final bag hanging from the setup during dwell five of five with fluid all over the floor, and felt that the leak was caused by the bag falling and ripping the line to drain line.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NI
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6075600
MDR Text Key59048746
Report Number1416980-2016-16901
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 11/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received10/10/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
HOMECHOICE
-
-