Model Number 16-02-80
Device Problems Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/11/2016
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Sorin group (b)(4) manufactures the sorin heater-cooler system 3t.The incident occurred in (b)(6).(b)(4).A sorin group field service representative was dispatched to the facility to investigate.The service representative replaced the pump motor on the patient bridge, which was found during troubleshooting to be the cause of the reported issue.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that patient circuit 2 of the sorin heater-cooler system 3t stopped working and the unit displayed an error code during a procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6075848
MDR Text Key59070464
Report Number9611109-2016-00782
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 10/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/20/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-