• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS HAND SWITCH F/EPD; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS HAND SWITCH F/EPD; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 05.001.012
Device Problem Failure to Shut Off (2939)
Patient Problem No Patient Involvement (2645)
Event Date 10/22/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).Device manufacture date is unknown.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This is report 2 of 3 of the same event.It was reported from india that the after plugging the electric pen drive device to the electric socket the device ran continuously.The electric pen drive device was in use with the cable device and hand switch device.The event did not occur during a surgical procedure.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device passed all manufacturing specifications.Therefore, the reported condition was not duplicated and confirmed.An assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAND SWITCH F/EPD
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
n/i
n/i
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
n/i
n/i
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6080150
MDR Text Key59511992
Report Number8030965-2016-15448
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.012
Device Lot NumberAV15676
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received11/10/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
CABLE DEVICE; ELECTRIC PEN DRIVE DEVICE
-
-