Model Number 302-20 |
Device Problems
Fluid/Blood Leak (1250); Cut In Material (2454)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/14/2016 |
Event Type
malfunction
|
Event Description
|
An implant card was received showing a full replacement for the patient during a prophylactic ifi replacement.Additional information was received from a company representative stating the reason the lead was replaced is the physician found the lead to have a cut in it during the replacement.The physician said that he did not cut the lead, therefore it seems to have been a pre-existing condition.He also said that the only pre-op diagnostic he saw was an ifi indicator.Per programming history date on (b)(6) 2016, diagnostics for the device were okay.Therefore the lead was likely functioning as intended up to the date of surgery and the patient was receiving intended therapy.The lead has not been received to date.
|
|
Event Description
|
The explanted lead was received on (b)(6) 2016.Analysis is underway, but hasn't been completed to date.
|
|
Event Description
|
Analysis was performed on the generator.The pulse generator performed according to functional specification.There were no performance or any other type of adverse conditions found with the pulse generator.
|
|
Event Description
|
Analysis was performed on the lead assembly.An abraded opening was identified in the outer silicone tubing.Note that since the electrode array portion was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Other than the above mentioned observations and typical wear and explant related observations, no other anomalies were identified in the returned lead portions.No obvious point of entrance was noted other than the identified outer tubing openings (for outer tubing fluid ingress) and the cut end of the returned lead portions (for inner tubing fluid ingress).
|
|
Search Alerts/Recalls
|