Model Number MODEL 250
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2016
Event Type  malfunction  
Event Description
During an internal review of programming and diagnostic history, it was identified that an unintended change in device settings occurred on (b)(6) 2016.There is no evidence of the settings been corrected on the same date; the patient most likely left the office visit with a duty cycle less than 2% (30 sec on time and 60 min off time).It was identified that an interrupted system diagnostic test occurred that could have caused an unintended change in device settings during that office visit on (b)(6) 2016.No patient adverse events were reported.
 
Event Description
Further information from the physician indicated that this patient's generator was replaced on (b)(6) 2016.Patient was then seen again in clinic in (b)(6) 2016 (which was the last visit until now), where his settings were adjusted.No further details on the settings could be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6089018
MDR Text Key59910840
Report Number1644487-2016-02585
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1063825
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received11/17/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/29/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-