The customer received questionable low elecsys afp assay results for one patient sample from cobas e602 analyzer serial number (b)(4).The results were 1.93 ng/ml, 6.00 ng/ml with a 1:10 dilution, and 2.02 ng/ml as the previous result for the patient had been about 10,000 ng/ml, the customer tested another sample from the patient drawn at the same time as the suspect sample and the result was about 10,000 ng/ml.Information concerning if any result was reported outside the laboratory was requested, but it was unknown.The patient was not adversely affected.
|
As part of the investigation, the sample from the patient that generated the result 2.02 ng/ml of was tested and the customer's result was reproduced.Elecsys: 1.83 ng/ml, lumipulse: 2.2 ng/ml.Based on the data available, a specific root cause could not be determined.The sample does not have a non-specific reaction or prozone phenomenon.Additional sample was not available for further testing.
|