• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. DISCOVERY ELBOW DISC ULNA 3X115MM LT W/BRNG C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. DISCOVERY ELBOW DISC ULNA 3X115MM LT W/BRNG C Back to Search Results
Catalog Number 114816
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Unspecified Infection (1930)
Event Date 10/10/2016
Event Type  Injury  
Event Description
Revision surgery - due to the patient having an infection, the surgeon replaced the ulna stem.
 
Manufacturer Narrative
The reason for this revision surgery was an infection. The previous surgery and the revision detailed in this investigation occurred 12 days apart. No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system. The healthcare professional indicated there was no delay in surgery and another suitable device was available for use. The revision surgery was completed as intended. The device was disposed of at the hospital and not made available to djo surgical for examination. A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements. There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection. The device was verified by biomet (zimmer) inc. To have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery. There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection. Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial). It is also possible that the patient was not compliant with post surgical instructions. There are multiple factors that may contribute to an infection that are outside of the control of djo surgical. There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDISCOVERY ELBOW
Type of DeviceDISC ULNA 3X115MM LT W/BRNG C
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
teffany hutto
9800 metric blvd.
austin, TX 78758-5445
5128346255
MDR Report Key6090756
MDR Text Key59551873
Report Number1644408-2016-00838
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K013042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Expiration Date10/02/2020
Device Catalogue Number114816
Device Lot Number333330
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/30/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/02/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 11/09/2016 Patient Sequence Number: 1
-
-