MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS COLIBRI INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS COLIBRI INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.001
Device Problems Moisture Damage (1405); Retraction Problem (1536)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/28/2016
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: (b)(4). (b)(6). The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment. Reliability engineering evaluated the device and the reported condition was confirmed. The assignable root cause was determined to be due to improper maintenance, which is user error/misuse/abuse. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported by (b)(6) that the reverse did not function on the small battery drive device. During service and evaluation, it was noted that there was liquid damage with the presence of water in the motor. It was not reported if the device was used in surgery, or if there was patient involvement. It was not reported if there were any delays in a surgical procedure or if a spare device was available for use. It was not reported if there were any injuries, medical intervention or prolonged hospitalization. The exact date of this event was unknown. All available information has been disclosed. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCOLIBRI
Type of DeviceINSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ CH4437
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6103086
MDR Text Key60436043
Report Number8030965-2016-15574
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 10/31/2016
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number532.001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2016
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/31/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-