Catalog Number PVPM |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/27/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
Attempts have been made to retrieve the device.
To date the device has not been returned.
If the device or further details are received at a later date, a supplemental medwatch will be sent.
|
|
Event Description
|
It was reported that the patient underwent an umbilical hernia repair procedure on (b)(6) 2016 and the mesh was implanted.
During the procedure, one of the wings of the mesh tore off when trying to place the mesh.
Another like mesh was used to complete the procedure.
There was no adverse patient consequence reported.
No further information is available.
|
|
Manufacturer Narrative
|
In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
The implant shows that one strap was broken off at the load-ring.
The cause of the strap breakage could not be verified.
The partial separation of the wings at the welding with the load ring may occur during handling and does not have any effect on the product properties and functionality.
It should be noted that before pvp placement enough space should be prepared to ensure proper unfolding of the device.
Once the mesh patch has been inserted into the defect, manipulate the suture loops to facilitate proper positioning of the patch.
|
|