Age at time of event: 18 years or older.Device is a combination product.(b)(4).Device evaluated by mfr: the stent delivery system (sds) was returned for analysis.A visual examination of the crimped stent found distal stent damage; the first distal strut was lifted and stretched in a proximal direction, struts on distal rows 2-4 were distorted.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination found no issue with the hypotube and with the shaft polymer extrusion profile.The bi-component bond showed no signs of damage or strain.A visual and tactile examination found no damage to the tip.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|