Model Number 302-20 |
Device Problem
High impedance (1291)
|
Patient Problem
No Code Available (3191)
|
Event Date 10/19/2016 |
Event Type
malfunction
|
Event Description
|
Prior to a generator replacement surgery on the date of surgery, a company representative detected high impedance (62;10,000 ohms).The patient's generator was being replaced due to ifi-yes.When the patient's new generator was implanted, high impedance was observed again.Therefore, the patient underwent lead revision as well and the impedance issue resolved.The suspect product was received.Product analysis has not be completed to date.No additional relevant information was received to date.
|
|
Event Description
|
The internal data of the generator was obtained from analysis of the device, and it was identified that the last >25% change in impedance occurred on (b)(6) 2016.No additional relevant information has been received to date.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
Product analysis was completed on the explanted lead.Two portions of the lead were returned, including the lead pin.The electrode array and a portion of the lead body were not returned.Based on the returned portions of lead analyzed, a lead fracture was not confirmed.The condition of the returned lead portions was consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no discontinuities were identified.No additional relevant information has been received to date.
|
|
Search Alerts/Recalls
|
|
|