The insulin pump passed displacement test, rewind, basic occlusion, occlusion, prime, and excessive no delivery alarm test.The pump functioned properly.The pump had intermittent button response noted during testing due to moisture damage on keypad traces.The pump was received with scratched lcd window, cracked reservoir tube lip, and cracked battery tube threads.(b)(4).
|
The customer reported via phone call that the pump had a keypad anomaly and high blood glucose.The blood glucose at the time of the incident was 176 mg/dl.The customer had symptoms such as tingling mouth, and nausea.The customer did not contact their healthcare professional, but do have a backup plan; manual injections.The drive support cap appeared normal.Troubleshooting was started, however the pump stopped responding.Troubleshooting was not able to resolve the issue.The device will be returned for analysis.
|