Model Number MODEL 250
Device Problems Crack (1135); Operating System Becomes Nonfunctional (2996)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/02/2016
Event Type  malfunction  
Event Description
It was reported that a physician was having issues with his programming system.The physician's handheld was frozen at the main screen and he was unable to make any selections.A hard reset was performed twice with no changes.The physician was unable to see if the pins on the flashcard were bent.There was also a reported crack on the handheld's screen that may have contributed to the issues.The handheld was reportedly sent to the manufacturer but has not been received to date.
 
Event Description
The tablet and the software were both returned and had product analysis completed.The software was verified to have no issues with responses on the touchscreen.The tablet cracked screen was verified and when it was replaced with a known good tablet, no further anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6121892
MDR Text Key60937225
Report Number1644487-2016-02712
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number180982
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received11/29/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/04/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-