The sample was not returned to davol for evaluation.
As reported during the implant procedure the balloon assembly was removed through a trocar and it was not noticed that the balloon had torn.
Regarding removal of the echo ps positioning system the instructions for use (ifu) state, "begin removal of the echo ps positioning system by grasping one of the two removal points marked by the dark arrows adjacent to the bard logo.
Begin pulling the positioning system off the mesh in one smooth motion.
Continue removing the echo ps positioning system, pulling it up to the tip of the trocar.
Remove both the echo ps positioning system and trocar simultaneously.
Verify that the echo ps positioning system including the balloon, all mesh connectors, and the inflation tube is fully intact, discard the echo ps positioning system appropriately.
Based on the events as reported the user did not identify the damage to the balloon assembly, which likely presented during attempted removal.
A review of the manufacturing records was performed and found that the lot was manufactured to specification.
The information provided by bard represents all of the known information at this time.
Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|