Model Number FUSION
Device Problem Imprecision (1307)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/27/2016
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.A medtronic representative inspected the imaging system on-site.The electromagnetic (em) control box and field generator were replaced and received by the manufacturer for evaluation.Analysis of the field generator could not duplicate the reported problem.The field generator was connected to a test system for an overnight burn-in test.The system remained in green status during all testing.It passed the accuracy test with an error of 1.806mm.Flexing the cable does not indicate any intermittent opens.Fully functional.Analysis of the box was similarly unable to duplicate the issue.The unit was connected to a test system with the ear, nose, and throat (ent) application for a burn-in test.Registration, tracking, and accuracy looked normal on all 4 tool ports.It passed the accuracy test with an error of 1.887mm.Connected / disconnected the tools from each port and system remained functional.Despite both parts being found fully functional, replacement of these parts resolved the issue.A full navigation system check-out was completed following part replacement and all tests passed.Full system functionality was confirmed and the system was returned to service.
 
Event Description
A medtronic representative reported that during an ear, nose, and throat procedure, the navigation system was 2mm inaccurate.The navigation system was still able to be used to successfully complete the procedure, and the patient was not impacted.No further details were provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
nicholas mcnabb
826 coal creek circle
louisville, CO 80027
7208902439
MDR Report Key6125230
MDR Text Key60877122
Report Number1723170-2016-05528
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 11/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received10/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-