• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problems Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/28/2016
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment and was able to replicate the issue.The medtronic representative found the eminterface showed one "flt" and by removing the patient reference frame from the axiem box solved the issue.After which the emitter and axiem communication issue status changed from red to green, flt disappeared.The medtronic representative opted to replace the emitter (field generator) after which the hardware, software, and instruments passed the system checkout.The system was fully functional.No parts have been received by manufacturer for analysis.
 
Event Description
A site representative reported that, while in a functional endoscopic sinus surgery (fess) the site was unable to track the patient tracker or instruments.The emitter details showed a communication error.The site rebooted the system to resolve the issue.The surgeon completed the procedure with the use of the navigation system.There was a reported delay to the procedure of 10 minutes due to this issue.There was no impact on patient outcome.
 
Manufacturer Narrative
Investigation of returned suspect field generator finds that the reported problem could not be duplicated.The field generator was connected to a test system for an overnight burn-in test.The system remained in green status during all testing.It passed the accuracy test with an error of 1.597mm.Flexing the cable does not indicate any intermittent opens.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
kristianna bilan
826 coal creek circle
louisville, CO 80027
7208902338
MDR Report Key6126354
MDR Text Key60877777
Report Number1723170-2016-05538
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age33 YR
Patient Weight37
-
-