• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number MERGE HEMODYNAMICS 10.0.3 PATCH 1
Device Problems Failure to Power Up (1476); Power Problem (3010); Noise, Audible (3273)
Patient Problem No Patient Involvement (2645)
Event Date 10/26/2016
Event Type  malfunction  
Manufacturer Narrative
The faulty ups was received by merge healthcare on 07nov2016 but has not yet been evaluated.For this reason, (b)(4) (conclusion not yet available-evaluation in progress) was used.When more information becomes available, a supplemental report will be submitted.
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that the ups (uninterrupted power supply) was buzzing and the hemo application was not powering on in the site's (b)(6).If parts of the hemo system become energized, there is a potential for direct harm to the patient and/or user including electrical shock or burns.However, no reports were received of staff injury and information obtained indicated that the problem occurred when no patient was present.Additionally, the site has other cath labs for use.Reference complaint number (b)(4).
 
Manufacturer Narrative
This supplemental report is submitted to the fda in accordance with the applicable regulations and as indicated by merge healthcare in the initial report submitted 23nov2016.The faulty ups (uninterrupted power supply) was returned to merge healthcare on (b)(6) 2016 for evaluation and was internally documented on (b)(6) 2017.The evaluation results showed that the unit powered on but would become very noisy at times for unknown reasons.Subsequently, the unit was scrapped by merge healthcare.No information was found in the tripp lite owner's manual for smart and omnismart medical grade ups systems (smart1200xlhg, (b)(4)) that addresses a buzzing issue.For this reason, conclusions code 57 (labeling deficiency) was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
MDR Report Key6126703
MDR Text Key60907227
Report Number2183926-2016-00802
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 10/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 10.0.3 PATCH 1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2016
Was the Report Sent to FDA? No
Date Manufacturer Received01/10/2017
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-