The complaint sample was visually inspected with no anomalies found.
To simulate the reported complaint condition, the valve was connected to heise gauge and tested at roller pump speed of 0.
5l/min, 1l/min and 2l/min.
At all speeds the valve opened at negative pressure of 149mmhg and positive pressure of 342mmhg.
The reported complaint condition could not be duplicated.
The device history records for the applicable lot of product were reviewed and the inspection report showed that no defect was found and rejected and no specific manufacturing yield issue was reported similar to this complaint condition.
Quest medical, inc.
Has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.
Quest medical, inc.
Defers to the patient's physician regarding medical history.
|