Device Problem
Temperature Problem (3022)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a temperature (temp - no physical damage) issue.There was no reported bodily injury due to the temperature issue.This complaint is being reported because there is the potential for the user to experience an injury to the skin due to increased pump temperature.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 02/01/2017 device evaluation: the device has been returned and evaluated by product analysis on 01/24/2017 with the following findings: a review of the black box showed no evidence of unusual voltage fluctuations.The rewind, load, and prime steps were performed successfully.The battery cap was undamaged and was able to fit securely.All currents read within specification.No overheating occurred during the investigation.The pump cover was removed to find no intermittent conditions to the printed circuit board or the continuous glucose monitoring module.The complaint of a temperature issue was unable to be duplicated.Unrelated to the original complaint the battery compartment was cracked below the grip pad.
|
|