MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/02/2016
Event Type  malfunction  
Manufacturer Narrative
Issue resolved at time of event.On 11/04/2016 a medtronic representative performed a navigation system check-out, all areas passed.System performed as intended.Fully functional.No parts have been received by manufacturer for analysis.Issue resolved at time of event.No further issues have been reported.
 
Event Description
A medtronic representative received a report from a distributor that while in a spine procedure, the site's navigation system screen became unresponsive within synergy spine 2.1.The mouse would not move.Performed a software re-boot which resolved the system issue.The navigation system was used for the remainder of the procedure without issue.Delay in therapy was less than one hour.There was no impact on patient outcome.
 
Manufacturer Narrative
The software investigation found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.System logs were analyzed by software engineering.Logs provided no additional insight regarding the unresponsive application.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
catherine eaton
826 coal creek circle
louisville, CO 80027-9710
7208902092
MDR Report Key6134094
MDR Text Key61094599
Report Number1723170-2016-02998
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 02/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? No
Date Manufacturer Received01/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/09/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-