• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problems Device Stops Intermittently (1599); Communication or Transmission Problem (2896)
Patient Problem No Patient Involvement (2645)
Event Date 11/04/2016
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Return requested.Replacement computer and camera both shipped to site 11/04/2016.Neither suspect device has been received by manufacturer for analysis.On 11/08/2016 a medtronic representative performed a navigation system check-out, software and instruments areas passed.Hardware test failed: optical communication.Replaced computer.Issue resolved.System performed as intended.
 
Event Description
A site representative reported that their navigation system camera was cycling and losing power every 4 minutes.No further details regarding the behavior, or specifically when it occurred, were provided.There was no patient present when this issue was identified.
 
Manufacturer Narrative
Analysis of suspect computer found it to be fully functional: system powered on and booted normally to login screen.App launched and navigated normally.No hdd or ram errors reported.Unable to replicate reported event.
 
Manufacturer Narrative
Correction: device manufacturing date now provided.A medtronic representative went to the site to test the equipment on 7/14/2017.The polaris spectra system control unit (scu) camera, positioning sensor unit (psu), and scu to psu cable were replaced.The navigation system passed the system checkout and was found to be fully functional.
 
Manufacturer Narrative
The positioning sensor unit (psu) for the navigation system was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.The polaris spectra system control unit (scu) for the navigation system was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.The electromagnetic interface box for the navigation system was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902092
MDR Report Key6135746
MDR Text Key61151158
Report Number1723170-2016-05609
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450968
UDI-Public00613994450968
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 02/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberS7
Device Catalogue Number9733858
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-