Model Number N/A
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Fall (1848)
Event Date 11/02/2016
Event Type  Injury  
Event Description
It has been reported that the patient underwent a total knee arthroplasty revision due to dislocation of the articular surface following a fall.
 
Manufacturer Narrative
Concomitant medical products: stem extension replacement screw, catalog # 00598009000, lot# 62778689; femoral component precoat for cemented use only size f left, catalog # 00599001601, lot# 63058138; tibial component stemmed precoat use of this tibial component with legacy knee - constrained condylar knee (lcck) articulating surfaces requires using, catalog # 00598003702, lot# 62899123; taper stem plug, catalog # 00596009900, lot# 62808326.The reported event was unable to be confirmed as not product was returned; visual and dimensional evaluations could not be performed.Dhr was reviewed and no discrepancies relevant to the reported event were found.Review of the complaint history determined that no further action is required as no trends were identified.Before this incident, patient hasn¿t experienced any pain.So, we can conclude that root cause is related to patient fall.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN LPS-FLEX ARTICULAR SURFACE
Type of Device
KNEE PROSTHESIS
Manufacturer (Section D)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer (Section G)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6136403
MDR Text Key61158518
Report Number0001822565-2016-04462
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
PK042271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Model NumberN/A
Device Catalogue Number00596203214
Device Lot Number62320985
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/24/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/11/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-