ZIMMER, INC. HGP II CUP FLEXIBLE DRILL; INSTRUMENT, HIP Back to Search Results
Model Number N/A
Device Problem Bent (1059)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/31/2016
Event Type  malfunction  
Event Description
It is reported that while drilling for the cup screws the drill bent due to the patient's hard bone.
 
Manufacturer Narrative
(b)(4).Complaint sample was evaluated and the reported event was not confirmed.The potential field service lifetime of the drill bit was 7.5 years.The flute of the drill bit is bent, rendering the instrument unusable.Dhr was reviewed and no discrepancies were found.Root cause could not be determined with information available, however it was mentioned that the patient had particularly hard bone which could have contributed to the damage.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no were trends identified.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HGP II CUP FLEXIBLE DRILL
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer (Section G)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6137673
MDR Text Key61195381
Report Number0001822565-2016-04468
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00661100302
Device Lot Number61296638
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-