Patient identifier and weight were not made available from the site.A medtronic representative, following-up at the site, reported the surgeons were performing an endoscopic procedure with trasnfenoidal approach using navigation.Surgeons realized that they were inaccurate when they were in the sellar floor (bony structures).The surgeon abandoned navigation and continued with the surgery based on his anatomical knowledge.The medtronic representative reported the same navigation system was used in a subsequent procedure successfully.There were no issues.- no further issues have been reported.Part not received by manufacturer.
|
A medtronic representative reported that, while in an ear, nose & throat (ent) transfenoidal procedure, two surgeons alleged an inaccuracy occurred during registration.The alleged inaccuracy was 2 millimeters anterior/posterior direction.No checkpoints were taken.The surgeon opted to discontinue the use of the navigation system to continue the procedure to completion.Delay in therapy was less than one hour.There was no impact on patient outcome.
|