Model Number S7
Device Problem Imprecision (1307)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/07/2016
Event Type  malfunction  
Manufacturer Narrative
Patient identifier and weight were not made available from the site.A medtronic representative, following-up at the site, reported the surgeons were performing an endoscopic procedure with trasnfenoidal approach using navigation.Surgeons realized that they were inaccurate when they were in the sellar floor (bony structures).The surgeon abandoned navigation and continued with the surgery based on his anatomical knowledge.The medtronic representative reported the same navigation system was used in a subsequent procedure successfully.There were no issues.- no further issues have been reported.Part not received by manufacturer.
 
Event Description
A medtronic representative reported that, while in an ear, nose & throat (ent) transfenoidal procedure, two surgeons alleged an inaccuracy occurred during registration.The alleged inaccuracy was 2 millimeters anterior/posterior direction.No checkpoints were taken.The surgeon opted to discontinue the use of the navigation system to continue the procedure to completion.Delay in therapy was less than one hour.There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
judith ericson
826 coal creek circle
louisville, CO 80027-9710
7208902187
MDR Report Key6139515
MDR Text Key61301039
Report Number1723170-2016-05619
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169505742
UDI-Public00643169505742
Combination Product (y/n)N
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733858
Is the Reporter a Health Professional? No
Date Manufacturer Received11/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age59 YR
-
-